Need Help? Contact us for technical support
Continue Shopping
TECHNICAL CHARACTERISTICS
Range
0 ÷ 20.00 ppm
Accuracy
± 1% (± 0.2mg/L if <5mg/L; ± 0.3mg/L if >5mg/L)
Response time
90% of the value in less than 60 s
Required flow
Not necessary
Temp. Compensation
NTC
Operating Temperature
0 ÷ 50 ° C
Storage Temp
-10 ÷ 60° C
Maximum Pressure
5 bar
Body Material
PVC
Electrode Material
Special optical glasses
O-Rings
NBR and Silicon
Luminophore cap diameter
10mm
Mechanical protection
IP68 sensor + cable
Mechanical Connection
3/4" BSP (Ø 36 mm)
Power Supply
12 ÷ 24 Vdc
Power Consuption
max. 2W
Electrical Connection
cable 10m included
Signal Interface
4-20mA